Color of Water
COLOR OF WATER
SPECIFICATIONS
Range 0 to 500 PCU (Platinum Cobalt Units)
Resolution 1 PCU
Accuracy ±10 PCU ±5% of reading
Typical EMC ± 1 PCU
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition
, Colorimetric Platinum Cobalt method.
REQUIRED ACCESSORIES
0.45 µm membrane for true color measurement.
For other accessories see page 73.
MEASUREMENT PROCEDURE
• Select the
Color of Water
method using the procedure described in
the
Method Selection
section (see page 12).
• Fill one cuvette up to the mark with deionized water and replace
the cap. This is the blank.
• Place the blank (# 1) into the holder and close the lid.
• Press ZERO key. The meter will show “-0.0-” when the meter is
zeroed and ready for measurement.
• Remove the blank.
10 mL
# 1
# 1
• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 6 minutes and press
READ. When the timer ends the meter will perform the reading.
The instrument displays concentration in µg/L of chromium VI.
• Press the or to access the second level of functions.
• Press the Chem Frm functional key to convert the result in µg/L of Chromate (CrO
4
2-
) and Dichromate
(Cr
2
O
7
2-
).
• Press the or to go back to the measurement screen.
INTERFERENCES
Interference may be caused by:
Vanadium above 1 ppm. However, waiting 10 minutes before reading, the interference is removed.
Iron above 1 ppm
Mercurous and mercuric ions cause slight inhibition of the reaction.
Chromium VI LR