INTERFERENCES
Interference may be caused by:
Co
2+
must not be present
Fe
2+
must not be present
Al
3+
above 32 mg/L
Ca
2+
above 1000 mg/L (as CaCO
3
)
Cd
2+
above 20 mg/L
Cl¯ above 8000 mg/L
Cr
3+
above 20 mg/L
Cr
6+
above 40 mg/L
Cu
2+
above 15 mg/L
F¯ above 20 mg/L
Fe
3+
above 10 mg/L
K
+
above 500 mg/L
Mg
2+
above 400 mg/L
Mn
2+
above 25 mg/L
Mo
6+
above 60 mg/L
Na
+
above 5000 mg/L
Pb
2+
above 20 mg/L
Zn
2+
above 30 mg/L
Nickel LR
Nickel LR
Press Timer and the display will show a countdown prior to adding reagent C or, alternatively, wait
for 15 minutes. Add one packet of HI 93740C-0 EDTA reagent to each beaker, cap and swirl to mix
until completely dissolved.
Fill one cuvette up to the mark with 10 mL of the blank.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Fill a second cuvette up to the mark with 10 mL of the reacted sample.
Insert the second cuvette into the instrument.
Press Read to start the reading. The instrument displays the results in mg/L of nickel.
Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking
readings, add 2-3 drops of Dispersing Agent (HI 93703-51) to each cuvette and swirl until turbidity
is removed.
10 mL
10 mL