DIRECT/USP MEASUREMENT
In this measure mode the user can check for ultra pure water using the United States Pharmacopeia standard
(USP <645>).
This USP standard consists of three stages (one in-line and two off-line tests) as following:
Stage 1 - this is an in-line test.
Test steps:
• Measure the temperature of the water and the uncompensating conductivity readings. The measurement
may be performed in a suitable container or as in-line measurement.
• The temperature will be round down to the nearest 5 °C and look up the corresponding conductivity value
in the below table.
• If the measured conductivity is lower than the conductivity in the table, then the water meets the USP
requirements.
• Otherwise, proceed to Stage 2 testing.
Stage 2 - this is an off-line test.
Test steps:
• Store the water sample in an enclosed clean container that has been rinsed previously with water of the
same quality.
• Adjust the sample’s temperature to 25 °C and agitate the sample to ensure that it has equilibrated with
ambient CO
2
.
• If the measured conductivity is less than 2.1 µS/cm, then the sample has met the USP requirements.
• Otherwise, proceed to Stage 3 testing.
erutarepmeT
)C°(
ytivitcudnoC
)mc/Sµ(
erutarepmeT
)C°(
ytivitcudnoC
)mc/Sµ(
06.0551.2
58.0062.2
019.0564.2
510.1075.2
021.1577.2
523.1087.2
034.1587.2
535.1097.2
047.1599.2
548.10011.3
059.1