Instruction Manual
SENSOR HANDLE
HI
4000-50
HI 4004HI 4004
HI 4004HI 4004
HI 4004
HI 4104HI 4104
HI 4104HI 4104
HI 4104
Calcium Ion
Selective Electrode
Half-cell
Combination
HI 4104
CALCIUM
COMBINATION
Vista de página 0
1 2 3 4 5 6 ... 15 16

Resumo do Conteúdo

Página 1 - Combination

1Instruction ManualSENSOR HANDLEHI4000-50HI 4004HI 4004HI 4004HI 4004HI 4004HI 4104HI 4104HI 4104HI 4104HI 4104Calcium IonSelective Electrode Half-

Página 2

107. Holding the body of the electrode gently press uppercap with your thumb. This permits the fill solution todrain out of the body. Release cap and

Página 3

11• Determine the difference between the two mV values.An acceptable value for this slope is 26 ± 4 mV (20-25°C).X. X. X. X. X. Corre

Página 4

122. Unscrew upper cap and slide down cable toward con-nector.3. Move spring and outer body down cable also.4. Dry off inner stem and module with a so

Página 5

13Procedure:Follow sections VIII and IX to prepare electrodes for mea-surement.1) Follow section VI to prepare standards/ solution. Stan-dards shoul

Página 6

14XII. XII. XII. XII. XII. Other Measurement TOther Measurement TOther Measurement TOther Measurement TOther Measurement Techniquesechniquesechniquese

Página 7

15pH range of 4-10 . The electrode responds to free calciumions only. Precipitates with oxalate, carbonate, phosphateand complexes with hydroxide,

Página 8

16MAN4104 07/06R1 WARRANTY WARRANTY WARRANTY WARRANTY WARRANTYHanna Instruments Ion Selective Electrodes are warrantedt

Página 9 - Deionized

2HI 4004 Calcium Half-cellHI 4004 Calcium Half-cellHI 4004 Calcium Half-cellHI 4004 Calcium Half-cellHI 4004 Calcium Half-cellHI 4104 Calcium Combinat

Página 10 - COMBINATION

3III. III. III. III. III. Theory of OperationTheory of OperationTheory of OperationTheory of OperationTheory of Operation:::::The HI 4004 and HI 4104

Página 11

4IV. IV. IV. IV. IV. Design elements of the HI 4004 and HI 4104Design elements of the HI 4004 and HI 4104Design elements of the HI 4004 and HI 41

Página 12

5V. V. V. V. V. Equipment required:Equipment required:Equipment required:Equipment required:Equipment required:• The HI 4004 requires the Ha

Página 13

6 • Due to shipping or storage the internal solution insidethe PVC modules may have developed an air pocketnear the membrane. Gently shaking the senso

Página 14

7Additional HI 4104 guidelines• Remove the protective plastic wrap that covers theceramic junction before assembling sensor for the firsttime.• Add re

Página 15

83. Holding the assembled electrode at the cable end,shake the sensor to ensure internal fill solution thatmay have separated during shipping is in co

Página 16 - MAN4104 07/06R1

94. Rinse inner stem with deionized water making certainto wet o-ring found on the inner stem.5. Reassemble electrode by gently pushing the innerassem