2
HI 4002 Bromide Half-cellHI 4002 Bromide Half-cell
HI 4002 Bromide Half-cellHI 4002 Bromide Half-cell
HI 4002 Bromide Half-cell
HI 4102 Bromide Combination ElectrodeHI 4102 Bromide Combination Electrode
HI 4102 Bromide Combination ElectrodeHI 4102 Bromide Combination Electrode
HI 4102 Bromide Combination Electrode
I. I.
I. I.
I.
Introduction:Introduction:
Introduction:Introduction:
Introduction:
The Hanna HI 4002and HI 4102 are ion selective elec-
trodes designed for the measurement of Bromide ions in
aqueous solutions. The HI 4002 is a solid state half-cell
sensor that requires a separate reference. The HI 4102 is a
combination ion selective electrode.
IIII
IIII
II.
SpecificationsSpecifications
SpecificationsSpecifications
Specifications
Type: Solid State electrode with
a Silver Bromide pellet.
Ion(s) measured: Bromide (Br
- -
- -
-
)
Measurement range: 1.0 M to 1 X 10
-6
M
79,910 to 0.08 ppm
Interfering ions: Cyanide, and Mercury
must be absent. Ratio of
interfering ion to Br
--
--
-
must
be less than the ratio
indicated below:
0.01 for I
--
--
-
iodide
0.01 for CN
--
--
-
cyanide
50 for Cl
--
--
-
Chloride
Operating Temperature: 0-80°C
Operating pH: 2-12.5 pH
Dimensions: 12 mm (OD) X 120 mm
nominal insertion
(0.47X 4.72)
Connection: BNC