EC calibration data (if available), which contains:
the number of calibrated standards (1 char)
the offset factor, with sign and decimal point
(7 chars)
the cell constant, with sign and decimal point
(7 chars)
the calibration time, yymmddhhmmss (12 chars)
standards information (for each standard)
• standard value, with sign and decimal point
(7 chars).
buffer unit (2 chars; 00-μS; 01-mS)
Reference Temperature with and decimal point
(4 chars)
Temperature Compensation mode (2 chars)
00 - no temperature compensation
01 - automatic temperature compensation
00 - manual temperature compensation
TC coeficient with sign and decimal point (4 chars)
calibration time, yymmddhhmmss (12 chars).
Na Cl Calibration data
• the number of calibrated data (1 char)
salinity coeficient, with sign and decimal point
(7 chars)
Cell constant, with sign and decimal point
(2 chars)
the calibration time (2 chars)
buffer information, for each buffer:
Reference Temperature with and decimal point
(4 chars)
Temperature Compensation mode (2 chars)
00 - no temperature compensation
01 - automatic temperature compensation
00 - manual temperature compensation
TC coeficient with sign and decimal point (4 chars)
calibration time, yymmddhhmmss (12 chars).
PAR Requests the setup parameters setting.
The answer string contains:
Instrument ID (4 chars)