Hanna-instruments HI 2550 Manual do Utilizador Página 47

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 46
47
Reading status (2 chars): R - in range, O - over
range, U - under range. First character corresponds
to the appropriate range reading. Second character
corresponds to mV reading (if exist).
Primary reading (corresponding to the selected range)
- 7 ASCII chars, including sign and decimal point.
Secondary reading (only when primary reading is not
mV, EC, NaCl, TDS) - 7 ASCII chars, including sign and
decimal point.
Temperature reading - 8 ASCII chars, with sign and
two decimal points, always in ºC.
MDR Requests the instrument model name and firmware code.
GLP Requests the calibration data record.
The answer string contains:
GLP status (1 char): represents a 4 bit hexadecimal
encoding.
0x01 - pH calibration available
0x02 - Rel mV calibration available
0x04 - EC calibration available
0x08 - NaCl calibration available
pH calibration data (if available), which contains:
the number of calibrated buffers (1 char)
the offset, with sign and decimal point (7 chars)
the average of slopes, with sign and decimal point
(7 chars)
the calibration time, yymmddhhmmss (12 chars)
buffers information (for each buffer)
type (1 char): 0 - standard, 1 - custom
status (1 char): N (new) - calibrated in last
calibration; O (old) - from an old calibration.
• warnings during calibration (2 chars): 00 - no
warning, 04 - Clean Electrode warning.
buffer value, with sign and decimal point
(7 chars).
the calibration time, yymmddhhmmss (12 chars).
Rel mV calibration data (if available), which contains:
the calibration offset, with sign (7 chars)
the calibration time, yymmddhhmmss (12 chars).