1Instruction ManualHI 250 HI 251Microprocessor-BasedpH/mV/ºCBench Meterswww.hannainst.com

Página 2 - WARRANTY

10pH CALIBRATIONCalibrate the instrument frequently, especially if high accuracy is required.The instrument should be recalibrated : Whenever the pH

Página 3 - GENERAL DESCRIPTION

11 The   tag will blink on the LCD until the reading isstable. When the reading is stable and close to the selected buffer,READY tag will be dis

Página 4 - FUNCTIONAL DESCRIPTION

12Note:  If the value measured by the instruments is not close to theselected buffer, WRONG BUF and WRONG   tags willblink alternately. In t

Página 5 - SPECIFICATIONS

13GOOD LABORATORY PRACTICE (GLP)GLP is a set of functions that allows storage and retrieval of data regarding themaintenance and status of the electro

Página 6

14 The calibration buffers in calibrating order.The first pH calibration buffer:The second pH calibration buffer:The third pH calibration buffer:Note

Página 7 - OPERATIONAL GUIDE

15LOGGING FUNCTIONUp to 50 LOG samples can be stored into memory.LOGGING THE CURRENT DATATo store the current reading into memory press LOG while the

Página 8

16VIEW LOGGED DATAPress RCL while in measurement mode to retrive the stored information.If no data were logged, the instruments display:Otherwise, the

Página 9

17LOGDATELOG mV The date on the primary LCD, along with DATE tag. The calibration offset on the primary LCD and OFS message on thesecondary LC

Página 10 - BUFFER pH

18SETUPSetup mode allows viewing and modifying the following parameters: Current Time Current Date pH Resolution Beep Status Baud Rate (serial co

Página 11

19CURRENT DATE SETPress CAL when the current date is displayed. The year will start blinking.Press the ARROW keys to change the year.Press RANGE. The

Página 12

2TABLE OF CONTENTSWARRANTY ... 2PRELIM

Página 13

20Press the ARROW keys to change the beep status (On or OFF).Press CFM to save the modified beep status.Press CAL to escape without saving.BAUD RATE S

Página 14

21INSTRUMENT ID PARAMETER SETPress CAL when the instrument ID is displayed. The instrument ID (0000 to9999) will start blinking.Press the ARROW keys t

Página 15 - LOGGING FUNCTION

22 Immerse the temperature probe in the second vessel as near as possible tothe reference thermometer. Allow a few seconds for the probe to stabilize

Página 16

23 Use the ARROW keys to set the reading on the secondary LCD to that ofthe hot water. When the reading is stable and close to the selectedcalibrati

Página 17

24RELATIVE mV CALIBRATION (HI 251)Note:  If the reading is not close to the selected calibration point,WRONG tag will blink. Verify calibration con

Página 18

25Data transmission from the instrument to the PC can be done with the HI92000 Windows® compatible software (optional). HI 92000 also offersgraphing a

Página 19

26CHR n Change the instrument range according with the parametervalue (n): n=1 pH range/0.01 resolution n=2 pH range/0.1 resolution n=3 mV range (H

Página 20

27The temperature has an effect on pH. The calibration buffer solutions areaffected by temperature changes to a lesser degree than normal solutions.Du

Página 21

28PREPARATION PROCEDURERemove the protective cap of the pH electrode.DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal withelectrodes. Th

Página 22 - TEMPERATURE CALIBRATION

29For refillable electrodes:If the filling solution (electrolyte) is more than 2½ cm (1) below the fill hole,add HI 7082 or HI 8082 3.5M KCl Electrol

Página 23 - mV CALIBRATION

3PRELIMINARY EXAMINATIONRemove the instrument from the packing material and examine it carefully tomake sure that no damage has occurred during shippi

Página 24

30CLEANING PROCEDURE General Soak in Hanna HI 7061 or HI 8061 GeneralCleaning Solution for approximately ½ hour. Protein Soak in Hanna HI 7073 or HI

Página 25 - PC INTERFACE

31TROUBLESHOOTING GUIDESMOTPMYS MELBORP NOITULOSevissecxe/esnoperwolS.tfird.edortceleHpytriD nipitedortceleehtkaoS1608IHro1607IHsetunim03rofnoituloseh

Página 26

32Typical Electrode LifeAmbient Temperature 1  3 years90 ºC Less than 4 months120 ºC Less than 1 monthAlkaline ErrorHigh concentrations of sodium ion

Página 27 - TEMPERATURES

33pH BUFFER SOLUTIONSHI70004P pH 4.01 Buffer Sachets, 20 mL, 25 pcsHI70007P pH 7.01 Buffer Sachets, 20 mL, 25 pcsHI70010P pH 10.01 Buffer Sachets,

Página 28 - & MAINTENANCE

34ORP PRETREATMENT SOLUTIONSHI7091L Reducing Pretreatment Solution, 460 mLHI7092L Oxidizing Pretreatment Solution, 460 mLpH ELECTRODESAll electrodes

Página 29

35HI1330BGlass-body, semimicro, single junction, refillable, combination pH electrode.Use: laboratory, vials.HI1331BGlass-body, semimicro, single ju

Página 30

36120 mm4.7"12 mm0.5"FC 10075 mm2.95"6 mm0.25"FC 200120 mm4.7"12 mm0.5"FC 210120 mm4.7"12 mm0.5"9.5mm DIA0.37&

Página 31 - TROUBLESHOOTING GUIDE

37Kynar® is registered Trademark of Pennwalt Corp.Ultem® is registered Trademark of General Electric Co.HI1413BGlass-body, single junction, flat

Página 32 - FOR pH SENSITIVE GLASS

38EXTENSION CABLE FOR SCREW-TYPE ELECTRODES(SCREW TO BNC ADAPTER)HI 7855/1 Extension cable 1 m (3.3') longHI 7855/3 Extension cable 3 m (9.9&ap

Página 33 - ACCESSORIES

39RECOMMENDATIONS FOR USERSBefore using these products, make sure they are entirely suitable for theenvironment in which they are used.Operation of th

Página 34 - 4.7"

41) Liquid Crystal Display (LCD).2) CFM/GLP key, to confirm different values or to display Good LaboratoryPractice information.3) CAL key, to enter an

Página 35 - "S" VERSION

40MAN25010/04SALES AND TECHNICAL SERVICE CONTACTSAustralia:Tel. (03) 9769.0666  Fax (03) 9769.0699China:Tel. (10) 88570068  Fax (10) 88570060Egypt:T

Página 36 - 0.25"

5HI 250SPECIFICATIONSEGNARHp0.61ot0.2Hp00.61ot00.2Cº0.021ot9.9NOITULOSERHp1.0Hp10.0Cº1.0YCARUCCAF°86/C°[email protected]±Hp10.0±Cº4.0±)rorreeborpgnidulcxe(e

Página 37 - ORP ELECTRODES

6HI 251SPECIFICATIONSEGNARHp0.61ot0.2Hp00.61ot00.2Vm9.996±Vm0002±Cº0.021ot9.9NOITULOSERHp1.0Hp10.0)Vm9.996±(Vm1.0)Vm0002±(Vm1Cº1.0YCARUCCAF°86/C°02

Página 38

7OPERATIONAL GUIDEPOWER CONNECTIONPlug the 12 VDC adapter into the power supply socket.Note: These instruments use non volatile memory to retain the p

Página 39

8If measurements are taken successively in different samples, it is recommendedto rinse the electrode thoroughly with deionized water or tap water and

Página 40

9RELATIVE mV MEASUREMENTS (HI 251 only) Press RANGE until rEL message will be displayed on the secondary LCDline for about one second and mV tag