Temperature has an effect on pH. The calibration buffer solutions are affected
by temperature changes to a lesser degree than normal solutions. During
calibration the instrument will automatically calibrate to the pH value corre-
sponding to the measured or set temperature.
TEMP pH VALU
ºC ºF 1.68 4.01 6.86 7.01
0321.67 4.01 6.98 7.13
5411.67 4.00 6.95 7.10
10 50 1.67 4.00 6.92 7.07
15 59 1.67 4.00 6.90 7.04
20 68 1.68 4.00 6.88 7.03
25 77 1.68 4.01 6.86 7.01
30 86 1.68 4.02 6.85 7.00
35 95 1.69 4.03 6.84 6.99
40 104 1.69 4.04 6.84 6.98
45 113 1.70 4.05 6.83 6.98
50 122 1.71 4.06 6.83 6.98
55 131 1.72 4.07 6.84 6.98
60 140 1.72 4.09 6.84 6.98
65 149 1.73 4.11 6.85 6.99
70 158 1.74 4.12 6.85 6.99
During calibration the instrument will display the buffer value at 25 °C.
pH VALUES AT DIFFERENT
TEMPERATURES