31
Rel mV calibration data (if available), which contains:
the calibration offset, with sign (7 chars)
the calibration time, yymmddhhmmss (12 chars).
PAR Requests the setup parameters setting.
The answer string contains:
Instrument ID (4 chars)
Calibration alarm time out (2 chars)
SETUP information (2 chars): 8 bit hexadecimal
encoding.
0x01 - beep ON (else OFF)
0x04 - degrees Celsius (else degrees Fahrenheit)
0x08 - Offset calibration (else Point calibration)
Auto-off/Light time (3 chars)
Auto power off time (3 chars)
The number of custom buffers (1 char)
The custom buffer values, with sign and decimal
point, for each defined custom buffer (7 chars)
NSLx Requests the number of logged samples (4 chars).
The command parameter (1 char):
P - request for pH range
M - request for mV and Rel mV ranges
LODPxxx Requests the xxx
th
pH record logged data.
LODMxxx Requests the xxx
th
mV/Rel mV record logged data.
LODPALL Requests all pH Log on demand.
LODMALL Requests all mV/Rel mV Log on demand.
The answer string for each record contains:
The logged mode (2 chars):
01 - pH range (0.01 resolution)
02 - pH range (0.1 resolution)
03 - mV range
04 - Rel mV range
Reading status (1 char): R, O, U
Calculated reading, with sign and decimal point
(7 chars) - for pH and Rel mV range
Temperature reading, with sign and two decimal
points (8 chars)
mV reading status (1 char): R, O, U - only for
HI 98150 and HI 98160