Press the ARROW keys to view the next
calibration parameter (pressing the key):
The pH calibration slope and offset.
The GLP slope is the average of the calibration slopes; the percentage is
referred to the ideal slope value. For example 59.16 mV/pH at 25 °C.
The calibration buffers in calibrating order,
for the last calibration.
The first pH calibration buffer:
The second pH calibration buffer (if exist):
The third pH calibration buffer (if exist):
The fourth pH calibration buffer (if exist):
The fifth pH calibration buffer (if exist):