Select the appropriate measurement range (EC or TDS).
Wait a couple of minutes for the temperature sensor to reach
thermal equilibrium. The display will then show the measurement
automatically temperature compensated for temperature with the
appropriate indication among the following:
µS symbol indicates the meter is in
EC mode;
ppm” symbol indicates the meter is
in TDS mode.
TAKING TEMPERATURE MEASUREMENTS
• Submerge the tip (4cm/1½") of the
probe into the sample to be tested.
Select the °C mode.
Stir briefly and wait a couple of
minutes for the reading to adjust
and stabilize. The display shows the
temperature value.
Notes: If the display shows a “1” on
the far left hand side and a
0” on the far right hand side,
the reading is out of range.
It is recommended to clean
often the probe with HI 700661
Cleaning Solution.
After measurements have been completed, the instrument
should be switched off, and the probe cleaned and covered
with the protective cap.