Página 1 - Zinc ISM

1Instruction ManualHI 96731Zinc ISMwww.hannainst.com

Página 2 - TABLE OF CONTENTS

10Under Range: A blinking “0.00” indicates that the sampleabsorbs less light than the zero reference. Check the procedureand make sure you use the sam

Página 3 - PRELIMINARY EXAMINATION

11The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts may c

Página 4 - ABBREVIATIONS

12• It is important that the sample does not contain any debris. This wouldcorrupt the reading.• Each time the cuvette is used, the cap must be tighte

Página 5 - PRECISION AND ACCURACY

13MEASUREMENT PROCEDURETo compensate the meter for the sample turbidity or color, the measurement takes place in twophases. First, the meter is zeroed

Página 6 - PRINCIPLE OF OPERATION

14• Remove the cuvette.• Add 0.5 mL of HI 93731B-0 Zinc ReagentB Cyclohexanone. This is the sample.Note: To prevent any contamination fromthe polycarb

Página 7 - FUNCTIONAL DESCRIPTION

15VALIDATION PROCEDUREUse the validation procedure to ensure that the instrument is properly calibrated.Warning: Do not validate the instrument with a

Página 8

16• Remove the cuvette.• Place the CAL CHECKTM StandardHI 96731-11 Cuvette B into the cuvetteholder and ensure that the notch on thecap is positioned

Página 9 - ERRORS AND WARNINGS

17CALIBRATION PROCEDURENote: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK orON/OFF keys.Warning: Do not calibra

Página 10

18• After a few seconds the display willshow “-0.0-”. The meter is now zeroedand ready for calibration. The blinking“READ” asks for reading calibratio

Página 11

19• Press GLP/ to edit the desired year(2000-2099). If the key is kept pressed,the year number is automaticallyincreased.• When the correct year has

Página 12 - (c)(a) (b)

2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DESC

Página 13 - MEASUREMENT PROCEDURE

20To display the calibration date:• Press GLP/ to enter GLP mode. Thecalibration month and day will appear onthe main display and the year on theseco

Página 14

21• Press READ/ /TIMER to enter in thefactory calibration restore screen. Theinstrument asks for confirmation of usercalibration delete.• Press ZERO/C

Página 15 - VALIDATION PROCEDURE

22• Turn the instrument off by pressing ON/OFF.• Turn the instrument upside down and remove the battery cover by turning it counterclockwise.• Extract

Página 16

23Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environmen

Página 17 - CALIBRATION PROCEDURE

24Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe

Página 18

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, pl

Página 19

4GENERAL DESCRIPTIONABBREVIATIONSThe HI 96731 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a m

Página 20 - GLP mode

5SPECIFICATIONSPRECISION AND ACCURACYRange 0.00 to 3.00 mg/LResolution 0.01 m/LAccuracy ±0.03 m/L ± 3% of reading @ 25°CTypical EMC Deviation ±0.01 m

Página 21 - BATTERY MANAGEMENT

6PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cross

Página 22 - BATTERY REPLACEMENT

71) GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cuvette h

Página 23 - WARRANTY

810852973461KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO/CFM: to zero the meter prior to measurement, to confirm edited values or t

Página 24 - MAN96731 09/10

9ERRORS AND WARNINGSThe instrument shows clear messages when erroneous condition appears. Messages are also displayedwhen the obtained values are outs