Página 1 - Bromine ISM

1www.hannainst.comInstruction ManualHI 96716CBromine ISM

Página 2 - TABLE OF CONTENTS

10Under range: A blinking “0.00” indicates that the sampleabsorbs less light than the zero reference. Check the procedureand make sure you use the sam

Página 3 - PRELIMINARY EXAMINATION

11The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amounts may c

Página 4 - ABBREVIATIONS

12MEASUREMENT PROCEDURE• Whenever the cuvette is placed into the measurement cell, itmust be dry outside, and completely free of fingerprints, oilor d

Página 5 - PRECISION AND ACCURACY

13• When the beeper sounds briefly and theLCD displays dashes, the meter is ready.The blinking “ZERO” indicates that theinstrument needs to be zeroed

Página 6 - PRINCIPLE OF OPERATION

14• Remove the cuvette.• Add the content of one packet ofHI 93716-0 Bromine reagent. Replacethe cap and shake gently for about 20seconds to dissolve m

Página 7 - FUNCTIONAL DESCRIPTION

15Alkalinity above 300 mg/L and acidity above 150 mg/L as CaCO3 require neutralization since thecolor could instantly fade. To resolve this, neutraliz

Página 8

16• After a few seconds, the display will show“-0.0-”. The meter is now zeroed and readyfor validation.• Remove the cuvette.• Place the CAL CHECKTM St

Página 9 - ERRORS AND WARNINGS

17Note: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK orON/OFF keys.Warning: Do not calibrate the instrument wit

Página 10

18• After a few seconds the display will show“-0.0-”. The meter is now zeroed and readyfor calibration. The blinking “READ” asks forreading calibratio

Página 11

19• Press GLP/ to edit the desired day(01-31). If the key is kept pressed, theday number is automatically increased.Note: It is possible to change th

Página 12 - MEASUREMENT PROCEDURE

2PRELIMINARY EXAMINATION ... 3GENERAL DESCRIPTION ...

Página 13

20In the GLP mode, the last user calibration date can be verified and the factory calibration can berestored.To display the calibration date:• Press G

Página 14

21• Press READ/ /TIMER to enter in thefactory calibration restore screen. Theinstrument asks for confirmation of usercalibration delete.• Press ZERO/C

Página 15 - VALIDATION PROCEDURE

22• Turn the instrument off by pressing ON/OFF.• Turn the instrument upside down and remove the battery cover by turning it counterclockwise.• Extract

Página 16

23Recommendations for UsersBefore using these products, make sure that they are entirely suitable for your specific application and for the environmen

Página 17 - CALIBRATION PROCEDURE

24MAN96716 12/10Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support

Página 18

3Please examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, please notify your Dealer

Página 19

4The HI 96716 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a manufacturer of analytical instru

Página 20 - GLP mode

5Range 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.08 mg/L ±3% of reading @ 25°CTypical EMC Deviation ±0.01 mg/LLight Source Tungsten lampLight

Página 21 - BATTERY MANAGEMENT

6Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam crosses a substance, some o

Página 22 - BATTERY REPLACEMENT

71) GLP/ key2) CAL CHECK key3) ZERO/CFM key4) READ/ /TIMER key5) ON/OFF key6) Liquid Crystal Display (LCD)7) Cuvette alignment indicator8) Cuvette h

Página 23 - WARRANTY

810852973461KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO/CFM: this is a bi-functional key. Just press to zero the meter prior to me

Página 24 - (ROMANIA)

9The instrument shows clear messages when erroneous condition appears. Messages are also displayedwhen the obtained values are outside expected range.