Instruction Manual
HI 93703
Portable Microprocessor
Turbidity Meter
Vista de página 0
1 2 3 4 5 6 ... 12 13

Resumo do Conteúdo

Página 1 - HI 93703

1Instruction ManualHI 93703Portable MicroprocessorTurbidity Meter

Página 2

1918Battery replacement must only take place in asafe area and using the battery types specifiedin this instruction manual.To install or replace theba

Página 3 - DETECTOR

2120* 1 FTU = 1 NTUBatteries are too weak toensure reliable measure-ments. This message ap-pears for a few seconds,then the meter will auto-matically

Página 4

2322CE DECLARATION OF CONFORMITYRecommendations for UsersBefore using this product, make sure that it is entirely suitable for the environmentin which

Página 5 - MEASUREMENT PROCEDURE

24MANTURBR4 08/05Australia:Tel. (03) 9769.0666 • Fax (03) 9769.0699China:Tel. (10) 88570068 • Fax (10) 88570060Egypt:Tel. & Fax (02) 2758.68

Página 6

32With HI 93703 turbidity measurements can beperformed with high precision in the field as wellas in the laboratory.HI 93703 turbidity meter is a port

Página 7 - SOURCES OF INTERFERENCE

54IR LEDEMITTED LIGHT (890 nm)CUVETLIGHTDETECTOR90° SCATTEREDLIGHTPRINCIPLE OF OPERATION* 1 FTU = 1 NTUHI 93703 has been designed to perform mea-surem

Página 8

76SPECIFICATIONS* 1 FTU = 1 NTURange 0.00 to 50.00 FTU *50 to 1000 FTU *Resolution 0.01 and 1 FTU *Accuracy ±0.5 FTU* or±5% of reading (whichever is

Página 9 - VIEWING THE CALIBRATION DATE

98• Wipe the cuvette thor-oughly with a lint-freetissue (HI 93703-70)before inserting intothe measurementcell. The cuvettemust be completelyfree of fi

Página 10

1110HOW TO ENSURE ACCURATE MEASURE-MENTS• Each time the cuvette is used, tighten thecap to the same degree.• Discard the sample soon after the reading

Página 11 - * 1 FTU = 1 NTU

1312CALIBRATIONA monthly calibration is recommended. Tocheck the date of last calibration, simply holdthe DATE/⇒ key down for few seconds.A more frequ

Página 12

1514• Insert the cuvette withthe HI 93703-10AMCO-AEPA-1 stan-dard @10 FTU* intothe measurement celland press the CAL keyagain.• After the second cali

Página 13 - MANTURBR4 08/05

1716STANDARD SUSPENSIONPresently, there are only two recognized pri-mary standards: AMCO-AEPA-1 and formazine.HANNA instruments supplies HI 93703 with