Hanna-instruments HI 901 Manual do Utilizador Página 104

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 148
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 103
8-4
AUXILIARY FUNCTIONS
To prime the burette, select Prime Burette from the Burette screen. Enter the number of
rinses and press .
The number of burette rinses can be set between 1 and 5 (we recommend at least three
rinses to assurre that the air bubbles are completely removed).
8.1.2 Rinse Tip
A 2 mL dose of titrant will be dispensed from the burette when this operation is started. This
operation will eliminate the air from the dispensing tip.
8.1.3 Manual Dispense
Using the Manual Dispense option, a defined titrant volume can be dosed. Select the Manual
Dispense option and press . The Manual Volume Dispense screen will become
active and the display will prompt you to enter the desired volume to be dispensed.
Vista de página 103
1 2 ... 99 100 101 102 103 104 105 106 107 108 109 ... 147 148

Comentários a estes Manuais

Sem comentários