Note: The reading of the first point could be skipped for FNU and NTU modes by pressing “Skip
functional key. In this case, the 0.00 value will be used for calibration.
Place the cuvette for first calibration point:
<0.1 FNU (NTU) calibration cuvette for all
modes except FAU mode.
15 NTU formazin prepared standard for FAU
mode.
Note: For the FAU mode, if necessary, press UP
or DOWN key to edit the calibration point value
to match the exact value of the standard.
Close the lid and press “Read” functional key. The display will show the value blinking and the
light icon during measurement.
Note: If the Average mode was previously selected, the measurement in calibration mode will be
done using the average.
At the end of the measurement, the second proposed calibration point is displayed.
SECOND POINT CALIBRATION
Remove the first calibration cuvette.
Place the cuvette for second calibration point:
15 NTU (FNU) calibration cuvette for NTU
ratiometric mode.
15 NTU formazin prepared standard for NTU
non-rating and FNU modes.
100 NTU formazin prepared standard for FAU mode.
Note:
If necessary, press UP or DOWN key to edit
each calibration point value to match the exact value of
the standard as measured with a reference turbidimeter.
Close the lid and press “Read” functional key.
The display will show the value blinking and the light icon during
measurement.
At the end of the measurement, the third proposed calibration point is displayed.
At this moment it is possible to exit calibration by pressing “End” functional key.
If the calibration is terminated, the display will briefly
show Store... and the two point calibration is saved.
The instrument returns to the main screen.
THIRD CALIBRATION POINT
Remove the second standard cuvette.
Place the cuvette for third calibration point:
100 FNU (NTU) calibration cuvette for NTU
ratiometric mode.
100 NTU formazin prepared standard for NTU
non-ratio and FNU modes.
750 NTU formazin prepared standard for FAU mode.
Note:
If necessary, press UP or DOWN key to edit each
calibration point value to match the exact value of the
standard as measured with a reference turbidimeter.
Close the lid and press “Read” functional key.
The display will show the value blinking and the light icon during
measurement.
At the end of the measurement, the fourth proposed calibration point is displayed.
At this moment it is possible to exit calibration by pressing “End” functional key.
If the calibration is terminated, the display will briefly
show Store... and the three point calibration is saved.
The instrument returns in the main screen.