Note:The reading of the first point colud be skipped by pressing the Skip key. In this case, the
0.00 NTU point will be used for calibration.
Place the <0.1 NTU standard cuvette (or the
cuvette with dilution water) into the holder and
ensure that the cuvette mark is aligned with the
sign on the instrument top.
Close the lid and press the Read key. The
display will show the value blinking and the lamp
icon during measurement.
Note:If the Average mode was previously selected, the measurement in calibration mode will be
done using the average.
At the end of the measurement, the second
proposed calibration point (15.0 NTU) is
displayed.
Remove the first standard cuvette.
Place the 15 NTU standard cuvette (or the second
formazyne prepared standard) into the holder, with
the cuvette mark aligned with the sign on the
instrument top.
Note:If necessary, press s or t key to edit
each calibration point value to match the
exact value of the standard as measured
with a reference turbidimeter.
Close the lid and press the Read key. The
display will show the value blinking and the lamp
icon during measurement.
If Non ratio Turbidity range is selected, the
display will briefly show “Store...” and the two
point calibration is saved. The instrument returns
in the main screen.
If Ratio Turbidity range is selected, at the end of
the measurement, the third proposed calibration
point (100 NTU) is displayed.
At this moment it is possible to exit calibration
by pressing the End key.
If the calibration is terminated, the display will
briefly show “Store...” and the two point
calibration is saved. The instrument returns in the
main screen.