Operational Guide
Procedure
A single one, two or three-point calibration can be performed, using the four
predened buers 4.01, 6.00, 8.30 and 10.01 pH. For a single point calibration
any of the four buers may be used, 8.30 pH is recommended.
Note: The HI 84529 will not accept other pH buers for calibration.
1. Pour small quantities of selected buer solutions into clean beakers. For
accurate calibration used two beakers for each buer solution, the rst
one for rinsing the electrode and the second one for calibration.
2. Put a magnetic stir bar in the beaker that will be used for calibration.
3. Remove the protective cap and rinse the electrode with some of the
buer solution to be use for the rst calibration point.
4. Put the rst beaker with calibration buer in the beaker holder.
5. Place the electrode holder on the top of the beaker and secure it by
turning clockwise and press STIR.
6. Immerse the pH electrode and the temperature probe approximately
2 cm (0.8”) into the buer, paying attention not to touch the stir bar.
To enter Electrode Calibration follow the next steps:
1. Press CAL function key then ELECTRODE.
2. The electrode calibration screen will be displayed.
3. Press CLEAR to delete the previous calibration.
1 Point Calibration
1. The 8.30 buer will be selected
by default. If necessary, press the
ARROW keys in order to select a
dierent buer value.
2. The (unstable measurement)
symbol will be shown on the display
until the reading becomes stable.
3. When the reading is stable and
close to the selected buer, the
(unstable measurement) symbol will
disappear and the CONFIRM key will
become active.
4. Press CONFIRM to conrm the
Electrode
Calibration
Procedure