Operational Guide
bu er value will be automatically selected.
14. After the second calibration point has been con rmed, press ESC to exit
without performing the third calibration point.
3 Point Calibration
1. Remove the electrode holder with
electrode from the top of the beaker.
2. Place the third beaker with
calibration bu er in the beaker
holder. Rinse the electrodes in a
beaker containing the third bu er
rinsing solution.
3. Place the electrode holder (with electrodes) on the top of the beaker
and secure it by turning clockwise and press STIR. If necessary press the
ARROW keys in order to select a di erent bu er value.
4. The (unstable measurement) symbol will be shown on the display until
the reading becomes stable.
5. When the reading is stable and close to the selected bu er, the
(unstable measurement) symbol will disappear and the CONFIRM key
will become active.
6. Press CONFIRM to con rm the calibration. The instrument stores the
calibration value and returns to calibration menu, where the date and
time for the pH calibration will be updated.
Note: A bu er con rmed during the calibration process is removed from
the list of available bu ers. If the value measured by the instrument is
not close to the selected bu er, a “Wrong Bu er” error message will
be shown on the display. Check if the correct bu er has been used or
regenerate the pH electrodes by following the Cleaning Procedure. If
necessary change the bu er or the electrode.
If the measured o set isn’t within the preset limits (±45 mV), the meter will
display the message “Bu er Contaminated” alternatively with ”Electrode
Dirty/Broken”.
Electrode
Calibration
Procedure