Hanna-instruments HI 84500 Manual do Utilizador Página 14

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 13
14
PUMP CALIBRATION MESSAGES
Prime burette screen when the dosing system is
running.
Pump calibration is initiated by pressing the Start
key.
This error message appears when the pump is not
working properly. Check the tubing, valve and
syringe. Press Restart to try again.
This error message appears during pump calibration
when the end point can not be reached and the
maximum amount of titrant is exceeded. Check
standard, electrode and/or dosing system and try
again.
This screen appears while pump calibration is in
progress. Press ESC or Stop key to return to the
Pump Calibration screen.
This screen appears when pump calibration is
complete.
Vista de página 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 35 36

Comentários a estes Manuais

Sem comentários