Hanna-instruments HI 84432 Manual do Utilizador Página 25

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 24
25
Press Purge to begin a purge cycle.
The purging stops automatically after 5 minutes.
To stop purging at any time and return to the main screen press ESC or Stop.
During a purge, a countdown timer displays remaining time in 5 minute purge.
Press Pause to interrupt the purge process.
Press Pause or Stop (by pressing the corresponding function key in the purge screen)
• after the first drops of fresh titrant appear at the dosing tip
• in case an error condition is observed (empty titrant bottle, tubes or dosing tip disconected,
pump error)
• if you want to resume at a later time
If Pause is pressed the next screen is displayed:
Press Resume to continue purging.
After the 5 minutes purging interval has elapsed the “Completed” message is displayed.
Another purge period can be initiated by pressing Restart or press ESC to return to main titrator
screen.
Vista de página 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 43 44

Comentários a estes Manuais

Sem comentários