1Instruction ManualHI 84430TOTAL ACIDITYMINITITRATOR & pH METERfor Water Analysiswww.hannainst.com

Página 2 - TABLE OF CONTENTS

1918ELECTRODE CALIBRATION PROCEDUREIt is recommended to calibrate the instrument frequently, especially if high accuracy is required.The pH electrode

Página 3 - GENERAL DESCRIPTION

2120• The calibrated value will be shown on the display and the third expected buffer value will beautomatically selected.• After the second calibrati

Página 4 - SPECIFICATIONS

2322• If the measured offset isn’t within the preset limits the meter will display the message “BufferContaminated” alternatively with ”Electrode Dirt

Página 5 - PRINCIPLE OF OPERATION

2524•Press CAL to access the titration calibration menu where electrode and pump calibration may be accessed.•Press HELP to view the contextual help.

Página 6 - TITRATOR STARTUP

2726• Press Pump.The next screen will be displayed.• Press Start.• After the pump calibration is started, on the upper right side of the display two

Página 7 - SETUP CONFIGURATION MENU

2928Verify: The instrument has been calibrated (pH and pump) before performing a lot of titrations.An electrode calibration in at least one point (8.3

Página 8

3130Note: If the end-point is not reached or it is not recognized or the input reading is out of range,an error message will be displayed. The titrati

Página 9

3332TITRATOR GLP INFORMATIONPress MENU while in Titration mode and then GLP.From this screen it is possible to select between viewing the electrode

Página 10 - ELECTRODE PREPARATION

3534Press LOG to memorize the current reading. A message indicating the free log space will bedisplayed for a few seconds.In order to take pH measure

Página 11

3736pH METER GLP INFORMATIONThe pH meter GLP screens references the last pH calibration data. GLP is an acronym for GoodLaboratory Practice.To view t

Página 12

32All rights are reserved. Reproduction in whole or in part is prohibited without the written consentof the copyright owner, Hanna Instruments Inc., W

Página 13 - PUMP TUBE INSTALLATION

3938TEMPERATURE CALIBRATION PROCEDURE (for technical personnel only)All the instruments are factory calibrated for temperature.HANNA’s temperature pro

Página 14 - PUMP CALIBRATION PROCEDURE

4140-Reading status (1 char): R-in range, O - over range, U - under range.-pH reading with sign and decimal point (7 chars)-Temperature, with sign and

Página 15 - TITRATION PROCEDURE

4342-Overrage for the measured value: R-in range, O-over range, U-under range.-Measured value (7 chars): for the pH meter log this is the measured pH

Página 16

4544ELECTRODE CONDITIONING AND MAINTENANCEpH CLEANING PROCEDURE• GeneralSoak in Hanna HI 7061 or HI 8061 General Cleaning Solution forapproximately ½

Página 17

4746HI 84430 is warranteed for two years against defects in workmanship and materials when used for itsintended purpose and maintained according to th

Página 18

48MAN84430R103/08SALES AND TECHNICAL SERVICE CONTACTSSALES AND TECHNICAL SERVICE CONTACTSSALES AND TECHNICAL SERVICE CONTACTSSALES AND TECHNICAL SERVI

Página 19

54The instrument employs a powerful and effective built-in algorithm to analyze the pH response todetermine the exact pH endpoint, then uses this to m

Página 20

76REQUIRED REAGENTSCode Description Quantity/TestHI 84430 - 50 Titrant (Low Range) 1 mLHI 84430 - 51 Titrant (High Range) 1 mLHI 84430 - 55M Pump Cali

Página 21 - PC INTERFACE

98PRINCIPLE OF OPERATIONWater acidity determination is based on the neutralization of all acidic species contained in watersamples by titration with b

Página 22

1110PERISTALTIC PUMPPeristaltic pumps are self priming. Liquid never contacts the pump components. The titrant tubingis pressed along the rotating rol

Página 23 - TROUBLESHOOTING GUIDE

1312From your knowledge of expected concentrations, use the table below to determine whichsettings are appropriate:Use Total LR or Strong LR for water

Página 24 - WARRANTY

1514Press the corresponding function key to change thelanguage.If the new selected language cannot be loaded, thepreviously selected language will be

Página 25 - MAN84430R1

1716Press Select to view the firmware version, languageversion, mV factory calibration date and time andtemperature factory calibration date and time.