The flashing digits indicate that:
1) the month can be set
2) the day can be set
“Storage”: this prompt appears for 1 second at the end of the
calibration procedure, to indicate that the calibration data has
been stored.
“Error”: the concentration of the calibration solution used is not
correct. Repeat the calibration procedure with the right standard
solution, and verify it is not expired. If the calibration procedure fails
again, contact your dealer or the nearest Hanna Customer Service
Center.
ERROR MESSAGES
a) on zero reading
This indicates that the zeroing procedure failed due to a low signal-
to-noise ratio. In this case press ZERO again.
“No Light”: the instrument cannot adjust the light level. Please check
that the sample does not contain any debris.
“Light Low”: there is not enough light to perform a measurement.
Please check the preparation of the zero cuvet.
“Light High”: there is too much light to perform a measurement.
Please check the preparation of the zero cuvet.
b) on sample reading
There is too much light for the sample measurement. Please check
if the right sample cuvet is inserted.
“Inverted”: the sample and the zero cuvet are inverted.
“Zero”: a zero reading was not taken. Follow the instruction in
the measurement procedure for zeroing the meter.